Indicated For:
|
Technical Data and Specifications |
|
Therapy Output Channels: |
Two Independent Therapy Channels |
|
Therapy Output Modes: |
True Interferential and Conventional Muscle Stimulation |
|
Waveform: |
Symmetrical biphasic square wave with zero net DC component |
|
Carrier Frequency: |
4000 Hz or 5000 Hz selectable |
|
Interference Frequency (IF): |
4001 Hz to 4150 Hz adjustable in 1 Hz increments OR
5001 Hz To 5150 Hz adjustable in 1 Hz increments. |
|
Net Therapeutic frequency: |
1 to 150 Hz - verifiable |
|
Output Voltage: |
25 volts peak / 0-50 volts peak-to-peak,
adjustable in 1/10th volt increment (.2ma eachincrement) |
|
Output Current: |
0-50 milliamps peak /0-100 milliamps peak to peak |
|
Pulse Width: |
125 micro seconds each phase |
|
Power System: |
2AA batteries OR EAZE UL approved ac wall adapter |
|
Wall Adapter: |
EAZE UL approved AC wall Adapter (included) |
|
Battery Charger: |
EAZE UL approved battery charger (included) |
|
Compliance Timer: |
Records Sessions (0-999) |
|
Dimensions: |
3.2" X 4.8" X 1.1" (81mm X 122mm X 27mm) |
|
Weight: |
6.2 oz. (176grams) including Batteries |
|
Preset Therapy Protocols: |
23 (11 Interferential Stimulation, 12 Muscle Stimulation,
6 of which are sequential stimulation protocols) |
Preset Therapy Protocols
Interferential Stimulation
(3) Abrupt frequency shifts
(2) Ramped frequency shifts
(2) Full Sweep frequency shifts
(3) Dual Therapy (Abrupt + Ramp)
Muscle Stimulation
(4) Conventional muscle stimulation protocols with varying contraction and relaxation cycle times
(2)Russian Stimulation (korts) protocols with 1000 Hz and 2500 Hz carrier frequencies
Sequential Stimulation
(3) Standard Sequence Muscle stimulation followed by interferential stimulation
(3)Standard Sequence Reverse Interferential stimulation followed by muscle stimulation
Special Safety and Convenience Features
-
Unique clinician selectable carrier frequency -4000 Hz or 5000 Hz - provides unprecedented alignment to clinical preference
-
Advanced microprocessors monitor device safety parameters and will shut unit down to preserve patient safety
-
Hardware and Software systems are constantly polling output levels to detect unsafe energy levels
-
Maximum therapy amplitude lock and intra-therapy amplitude lock protects patient from unwanted amplitude levels
-
Pause & resume allows immediate cessation of therapy-therapy intensity upon resumption is patient controlled
-
Therapy lockout (set by healthcare provider or dealer) prevents patient from accessing non-beneficial therapies
-
Selectable operating language in English, Spanish and French - leads to greater patient undertanding and compliance
-
Up-to-date, sophisticated electronic circuitry utilizing high quality components assures maximum patient comfort
-
The Ultimate Ortho-Relief 2 is powered by either readily available and easily replaced AA batteries or the included AC adapter
-
The Ultimate Ortho-Relief 2 is supplied with (4) Nickel Metal Hydride rechargeable batteries and a UL approved charger